3 • Operation and Maintenance

Doc. 065291-01 3/09

65

3.10.4 Injecting Samples

After loading the sample in the sample loop, use one of the following
methods to switch the injection valve to the Inject position:

Manually: Click the

Inject

button on the Chromeleon Control panel

(see

Figure 3-4

) or touch the

Inject

button on the

HOME

page.

Automatically: Include the Inject command in a Chromeleon
program. (If necessary, refer to the Chromeleon Help or user manual
for assistance.)

3.11 Processing Samples

Samples can be run manually (one at a time) or else they can be grouped and run
automatically in batches.

3.11.1 Manual Sample Processing

To process samples manually, select operating parameters and commands
from the touch screen pages, or from the Chromeleon menu bar, toolbars,
and Control panel. Commands are executed as soon as they are entered.

NOTE When you run samples manually from the front

panel (without using

Chromeleon), the data is saved

only temporarily in the ICS-2100 memory. For a
permanent record of the data, connect a recording
device (chart recorder, integrator, etc.) to the analog
output. See

Section B.21

for analog output

connection instructions.

1. Complete the instructions in

Section 3.2

through

Section 3.8

to

prepare the ICS-2100 for operation and to prepare the sample for
processing.

2. Load the sample, using a syringe, vacuum syringe, or autosampler

(see

Section 3.10

).

3. In Chromeleon, select Control > Acquisition On. Chromeleon

records the signal supplied by the detector and displays the signal plot
on the Control panel.