2.1.2

Top Cover

Figure 2-7

illustrates the top cover of the ICS-2100.

EluGen Cartridge and Reservoir
Storage

The top cover holds one or two
EluGen

®

cartridges and up to

two 2-L plastic reservoirs
(P/N 044129) or one 4-L plastic
reservoir (P/N 039164).

EluGen Cartridges

The EluGen cartridges are
installed in holders that fit into
the two back reservoir areas.
See

Section 2.4.4

for details

about the eluent generator.

Tubing Chase

The tubing chase under the
EGC and CR-TC connectors
routes tubing from the eluent
reservoir and EGC holder to the
front of the ICS-2100.

Figure 2-7. ICS-2100 (Top View)

EGC, EPM, CR-TC, and Auxiliary Power Supply Connectors

The cable from the EluGen cartridge (EGC) connects to the

EGC-1

connector. The cable from the Continuously Regenerated Trap Column
(CR-TC) connects to the

CR-TC

connector. If a second EluGen cartridge is

installed, it connects to the

EGC-2

connector. If the second eluent

generator power supply is configured to power an auxiliary electrolytic
device, it connects to the

EGC-2

connector. For details, refer to the ICS-

2100 installation instructions.

When operating with an EGC II K

2

CO

3

EluGen Cartridge and the EPM

Electrolytic pH Modifier, the EluGen cartridge is typically connected to
the

EGC-1

connector. The pH modifier, which mounts on the side of the

Deionized
Water

EluGen
Cartridge
and Holder

EGC, EPM,

CR-TC, and

Auxiliary

Power

Supply

Connectors

Tubing

Chase

(under

connectors)

EGC Service

Area

4 L
Reservoir
Area

Second
EluGen
Cartridge or
Reservoir
Area