5 • Service

Doc. 065291-01 3/09

105

let it stabilize. When the temperature has stabilized, “=” is displayed
to the right of the temperature reading on the

CONDUCTIVITY CELL

page.

11. Select

1.00 mL/min

to begin pumping 1.00 mM KCl through the cell

at 1.0 mL/min.

12. Wait until the total conductivity reading stabilizes (in about

15 minutes) and then select

Calibrate

on the Wellness Panel or the

touch screen

CONDUCTIVITY CELL

page.

After calibration, the conductivity reading should be 147.00 ± 2

μ

S

and the cell constant should be between 120 and 180. If this is not the
case, contact Dionex for help.

13. If you are using the Wellness Panel, click Log to record the new

calibration values in the Audit Trail.

14. If you are using the touch screen, record the new calibration values in

the Chromeleon Wellness database. To do this, open the Wellness
Panel (see

Section 5.1.1

) and click Upload under Update Wellness

Database.

15. Flush the KCl solution from the system by pumping deionized water

through the cell. When the conductivity drops to less than 1

μ

S/cm,

stop the pump.

16. Reconnect the pump to the injection valve and reconnect the line

from the suppressor

ELUENT OUT

port to the cell inlet.

17. Reconnect the cell outlet to the suppressor

REGEN IN

port.