5.17 Priming the Pump with Isopropyl Alcohol

NOTE Prime the pump heads with isopropyl alcohol only if the

two standard priming procedures described in

Section 5.16

are unsuccessful.

1. Connect a 10-mL syringe

(P/N 054578) filled with
isopropyl alcohol (IPA) to
the port in the primary
pump head (see

Figure 5-33

).

2. Open the waste valve on

the secondary pump head
(see

Figure 5-33

) by

turning the knob one-
quarter to one-half turn
counterclockwise.

Figure 5-33. Priming the Pump Heads with IPA

3. In the Pump Settings window (see

Figure 3-6

), click Pump On.

4. Open the priming valve on the primary pump head by turning it one-quarter to

one-half turn counterclockwise.

5. Use the syringe to slowly push alcohol through the pump.

NOTE Be careful not to push any air trapped in the syringe

through the pump. Check the waste line from the
secondary pump head to verify that there are no air
bubbles.

6. Close the priming valve. Do not overtighten. Disconnect the syringe from

the priming valve.

7. Let the pump run for several minutes, to purge alcohol from the pump heads.

Then, click the Prime button in the Pump Settings window to flush the heads
with the desired eluent.

10 mL Syringe

(P/N 054578)

Priming Valve

(open)

Port

Waste Valve

(open)

Primary
Pump Head

Secondary

Pump Head