You can select the concentration of eluent to be generated from either the
Chromeleon Control panel or from the touch screen

HOME

page (see

Section B.4

).

The allowable eluent concentration depends on several factors: the flow
rate, suppressor type, EluGen cartridge type, and cartridge configuration.
For details, refer to

Table 2-2

and

Table 2-3

.

Single-Cartridge Configuration

In the single-cartridge configuration, the ICS-2100 contains one EluGen
cartridge.

Linked Dual-Cartridge Configuration

In the linked dual-cartridge configuration, the ICS-2100 contains two
EluGen cartridges (or one EluGen cartridge and an EPM) that perform as
one cartridge. Note that the allowable eluent concentration for a linked
cartridge is less than when the cartridge is defined as independent. For

EluGen Cartridge

Eluent Concentration Range

K

2

CO

3

0.1 to 15 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 2.0 mL/min flow
where X = 15/flow in mL/min

KOH

0.1 to 100 mM at

0.1 to 1.0 mL/min flow

0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min

LiOH

0.1 to 80 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 80/flow in mL/min

MSA

0.1 to 100 mM at

0.1 to 1.0 mL/min flow

0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min

NaOH

0.1 to 100 mM at

0.1 to 1.0 mL/min flow

0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow mL/min

Table 2-2. Eluent Concentration Ranges for Single-Cartridge

Configurations