7 set system operating conditions, Set system operating conditions – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 70

Advertising
background image

ICS-2100 Ion Chromatography System

58

Doc. 065291-01 3/09

reminder message appears. The pump will begin pumping at approximately
3 mL/min.

4. Continue priming the ICS-2100 until all air and previous eluent are purged

and no air bubbles are exiting the waste line.

5. Press

Pump

Off

.

6. Close the waste valve. Do not overtighten. The pump is now ready for

operation.

3.7

Set System Operating Conditions

NOTE This section is an overview of the steps needed to start

up the system and begin running samples. Actual
operating parameters (flow rate, cell heater
temperature, suppressor current, etc.) depend on the
application to be run. Refer to the column manual for
the required parameters for your application.

Set or verify system operating parameters from either the Chromeleon

Control

panel or from the touch screen

HOME

page. Operating parameters can also be set

automatically by loading a Chromeleon sequence.

NOTE Clicking the System Startup button on the Chromeleon

Control panel starts the pump and suppressor. The flow
rate and suppressor current settings that were in effect
when the system was shut down are restored.

NOTE Clicking the System Startup button on the Chromeleon

Control panel starts the pump, suppressor, eluent
generator, and CR-TC. The flow rate, suppressor
current, and eluent concentration settings that were in
effect when the system was shut down are restored.

1. Verify that the pump is on and set to the correct flow rate.

2. Verify that the suppressor current is on and that the setting is correct.

3. Verify that the eluent generator is turned on and the correct concentration is

set. Verify that the CR-TC is turned on.

4. Verify that the cell heater is set to the correct value.