3 • Operation and Maintenance

Doc. 065291-01 3/09

51

3.2

Turning On the System Power

Press the power switch on the ICS-
2100 rear panel (see

Figure 3-3

) to

turn on the system power.

Table 3-4

shows the ICS-2100 conditions at
power-up.

Also turn on the power to the
computer and the autosampler (if
used).

Figure 3-3. ICS-2100 Rear Panel

Feature

Power-Up Condition

Pump

Off

Injection valve

Load position

Cell

Reading current value

Suppressor

Off*

Cell heater

Set to the last value used. The default when the ICS-2100
is turned on for the first time is 35 °C.

Column oven temperature

Set to the last value used. The default when the ICS-2100
is turned on for the first time is 30 °C.

Eluent generator

Off*

CR-TC

Off

EPM

Off

*

* When you start the suppressor, eluent generator, or EPM, the value used last is restored.

Table 3-4. ICS-2100 Power-Up Conditions