5 auxiliary power supply (optional), 6 injection valve, Auxiliary power supply (optional) – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 54: Injection valve, Section 2.4.6, Ics-2100 ion chromatography system

Advertising
background image

ICS-2100 Ion Chromatography System

42

Doc. 065291-01 3/09

the EPM to achieve the desired 3.50 mM K

2

CO

3

/1.00 mM KHCO

3

mixture.

2.4.5

Auxiliary Power Supply (Optional)

The EGC-2 power supply can be configured as an auxiliary power supply
for the operation of an electrolytic device such as a water purifier. The
power supply operates in constant current mode and can be configured
from 0 to 200 mA, with a maximum voltage of 35 V.

2.4.6

Injection Valve

The injection valve (P/N 057968) is a six-port, electrically-activated
valve. A 25-

μL sample loop (P/N 042857) is installed on the valve at the

factory.

The valve has two operating positions: Load and Inject (see

Figure 2-18

).

Eluent flows through either the Load or Inject path, depending on the
valve position.

In the Load position, sample is loaded into the sample loop, where it
is held until injection. Eluent flows from the pump, through the valve,
and to the column, bypassing the sample loop. Sample flows from the

Figure 2-18. Injection Valve Flow Schematics

To Waste

Sample In

To Column

From Pump

Sample
Loop

LOAD POSITION

To Waste

Sample In

From Pump

Sample
Loop

INJECT POSITION

To Column