C • TTL and Relay Control

C.1

TTL and Relay Connections

A 12-pin connector strip for TTL/relay control is located on the ICS-2100 rear
panel. The connector provides two relay outputs, two TTL outputs, and four TTL
inputs (see

Figure C-1

).

NOTE TTL 1 Input - Inject/Load applies only to the injection

valve (not to the auxiliary valve, if installed).

Figure C-1. TTL and Relay Connector on Rear Panel

Relay loads in excess of 200 mA or with included power supplies over
60 V may damage the relay drivers on the CPU.

1

2

OUT

RELAY

(+)

TTL OUT

1

2

TTL GND

(-)

1

2

3

4

TTL IN

(+)

1

2

1

2

Description

Solid State Relay Contacts Output

Pin Function

3

4

Solid State Relay Contacts Output

TTL Output 1 (1 k pull up to +5, 100 mA sink)

Ω

5

TTL Output 2 (1 k pull up to +5, 100 mA sink)

Ω

6

7

8

9

10

11

12

Ground

Ground

TTL Input 2

⎯ Autozero

TTL Input 3

Pump/Suppressor

On

TTL Input 4

Mark

Note: The TTL input
functions can be reassigned
to different inputs.

Connector

Position

TTL Input 1

⎯ Inject/Load