A.7 detector electronics, A.8 conductivity cell with heat exchanger, Detector electronics – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 212: Conductivity cell with heat exchanger, A.7 detector

Advertising
background image

ICS-2100 Ion Chromatography System

200

Doc. 065291-01 3/09

A.7

Detector

Electronics

A.8

Conductivity Cell with Heat Exchanger

Type

Microprocessor-controlled digital signal processor

Cell Drive

8 kHz square wave

Detector

Linearity

1% up to 1 mS

Resolution

0.00238 nS/cm

Full-Scale

Output Ranges

Digital signal range: 0 to 15,000

μS

Analog signal range: 0 to 15,000

μS

Electronic Noise

±0.1 nS/cm when background conductivity is 0 to 150

μS/cm

±2 nS/cm when background conductivity is 151 to 3200

μS/cm

Filter

Rise times from 0 to 10 s; user-selectable

Temperature

Compensation

Fixed at 1.7% per 1 °C at cell temperature

Temperature

Range

Ambient +7 °C; 30 to 55 °C

Temperature

Stability

<0.01 °C

Cell Body

Chemically inert polymeric material

Cell Electrodes

Passivated 316 stainless steel; compatible with MSA

Cell Volume

<1

μL

Maximum Cell

Operating

Pressure

10 MPa (1500 psi)

Heat Exchanger

Inert, tortuous path for low axial dispersion