ICS-2100 Ion Chromatography System

86

Doc. 065291-01 3/09

4.3

Liquid Leaks

Leaking fitting

Locate the source of the leak. Tighten or, if necessary, replace the liquid line
connection (see

Section 5.3

). Refer to Installation of Dionex Liquid Line

Fittings (Document No. 031432) for tightening requirements. The manual is
included on the Dionex Reference Library CD-ROM (P/N 053891).

Broken liquid line

Replace the line and fittings with the same length and internal diameter tubing
(see

Section 5.3

).

Blocked or improperly installed line

Make sure the lines are not crimped or otherwise blocked. Also, if the blocked
line is a waste line, make sure it is not elevated at any point after it exits the
ICS-2100. If a line is blocked, replace it (see

Section 5.2

).

Loose pump check valve

1. Make sure the check valves are firmly seated in the pump head. If they are

not, tighten them carefully with an open-end wrench just until the leak
stops.

2. If the leak persists, replace the check valve (see

Section 5.6

).

Leaking seal wash port

Leaking from the seal wash port indicates a leaking piston seal. Replace the
seal (see

Section 5.7

).

Damaged pump piston seal

1. Replace the piston seal (see

Section 5.7

).

2. If the leak persists, replace the piston (see

Section 5.8

).

Pump head not tight against casting

Carefully tighten the pump head mounting nuts just until the leak stops. DO
NOT OVERTIGHTEN!