14 replacing the eluent valve – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 147

Advertising
background image

5 • Service

Doc. 065291-01 3/09

135

tubing) and press the heat exchanger onto the heater. Replace the screws
removed in

Step 8

.

10. Connect the new heat exchanger inlet line to port C on the injection valve and

connect the outlet line to the guard column inlet.

11. Reinstall the top metal plate and the heater cover.

5.14 Replacing the Eluent Valve

1. Turn off the power to the ICS-2100.

2. Open the ICS-2100 front door.

3. To prevent an eluent leak during the valve replacement procedure, disconnect

the eluent line from the reservoir cap or plug the line on the left side of the
valve with a coupler (P/N 039056) and a plug (P/N 037268) after you
disconnect the liquid line from the valve in

Step 4

.

4. Disconnect the two liquid

lines connected to the
eluent valve (see

Figure 5-26

).

5. Unscrew and remove the

two screws that attach the
valve to the component
panel (see

Figure 5-26

).

Save the screws.

Figure 5-26. Eluent Valve

Remove

Disconnect