5.10 Replacing the Conductivity Cell

1. Turn off the ICS-2100 power.

2. Open the front door and disconnect the tubing from the

CELL IN

and

CELL

OUT

fittings (see

Figure 5-18

, View A). Loosen the two screws on the cell

front cover. The screws remain attached (see

Figure 5-18

, View B).

Figure 5-18. Removing the Cell Front Cover

3. Grip the two loosened screws

and pull the cell cover toward
you to remove it from the
component panel. Do not pull
from the

CELL IN

and

CELL

OUT

fittings.

4. Continue pulling on the cover

until the attached cell assembly
slides out the front of the ICS-
2100 (see

Figure 5-19

).

Figure 5-19. Removing the Cell Assembly

Disconnect

Unscrew

Unscrew

View A

View B