5 • Service

Doc. 065291-01 3/09

131

10. Push the cables into the opening in the component panel.

11. Align the top of the suppressor

with the guide line printed on the
component panel (see

Figure 5-22

) and align the slots

on the rear of the suppressor
with the tabs on the panel.

12. Press the suppressor onto the

tabs and then slide it down until
it locks into place. Pull slightly
on the center of the suppressor to
verify that it is securely fastened.
Note: Some suppressors require
more force to secure them onto
the tabs.

13. Connect the two eluent and two

regenerant lines to the new
suppressor.

14. Close the ICS-2100 front door.

Figure 5-22. Suppressor Guidelines and Mounting

Tabs on Component Panel

5.12 Replacing the Column Heater

1. Turn off the ICS-2100 power.

2. Open the front door of the ICS-2100.

3. Remove the existing column heater:

a.

Unscrew the two thumbscrews on the heater cover (they remain attached
to the cover).

b. Pull the heater cover straight out to remove it.

c.

Unscrew the six thumbscrews on the top metal plate (see

Figure 5-23

) and

remove the plate.

1

1

2

3

4

1

Suppressor
Mounting Tabs

Suppressor
Cable Connector

Guideline for
AES

Guideline for
SRS

1

2

3

4