Ics-2100 ion chromatography system – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 184

Advertising
background image

ICS-2100 Ion Chromatography System

172

Doc. 065291-01 3/09

1. Disconnect the

TO INJ VALVE IN - P/ELUENT OUT

line from the EGC

CO

3

Mixer inlet.

2. Use a 10-32 to 10-32 coupler (P/N 042627) to connect the yellow

0.5 mL/min, 7 MPa (1000 psi) backpressure coil (P/N 053765) to the

TO INJ VALVE IN - P/ELUENT OUT

line.

3. Connect a 10-32 to 1/4-28 coupler (P/N 042806) to the free end of the

backpressure coil.

4. Disconnect the

TO SRS/AES REGEN OUT

line from the suppressor

REGEN OUT

port.

5. Connect the

TO SRS/AES REGEN OUT

line to the 10-32 to 1/4-28

coupler (see

Figure 5-66

).

NOTE The backpressure coil connection in

Figure 5-66

is

temporary; it is used only during the hydration and
conditioning procedure described in the next
section.

Figure 5-66. Backpressure Coil Connections for Hydrating and

Conditioning the EPM with Carbonate Eluent

10-32 to

1/4-28

coupler

TO INJ
VALVE
IN - P/
ELUENT OUT

Backpressure
Coil

TO SRS/AES

REGEN OUT

10-32 to
10-32 coupler