5 reinstalling the elugen cartridge – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 205

Advertising
background image

5 • Service

Doc. 065291-01 3/09

193

Connect the line labeled

TO INJ VALVE IN - P/ELUENT OUT

to the

EGC CO

3

Mixer inlet.

Connect the line labeled

TO PUMP/DAMPER

to the pulse damper.

4.

Direct the

WASTE, GAS SEPARATOR

line from the top of the ICS-

2100 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see

Section B.10.1

).

5.

Go on to

Section 5.23.5

.

5.23.5 Reinstalling the EluGen Cartridge

1. Pull the coiled black

EGC IN

and

EGC OUT

lines out from inside the

EGC holder. Disconnect the

EGC IN

line from the coupler that

connects it to the

EGC OUT

line.

2. Connect the

EGC IN

line

to the EluGen

cartridge

INLET

port

(see

Figure 5-82

).

3. Remove the coupler

and connect the

EGC

OUT

line to the

EluGen cartridge

OUTLET

port.

Figure 5-82. EluGen Cartridge Inlet and Outlet Port Connections

4. Turn over the EluGen cartridge (fittings facing down). Check for any

bubbles in the eluent generation chamber, and if necessary, shake and
tap the EluGen cartridge to remove them.

5. Reinstall the EluGen cartridge in the holder.

6. Reconnect the EluGen cartridge cable and the CR-TC or EPM cable

(if installed).

Remove the cap from the luer adapter on the EluGen cartridge vent
opening and re-connect the

EGC VENT

line.

EGC OUT

OUTLET
Port

EGC IN

INLET

Port