4.13 High Cell Output

Background not suppressed by suppressor

Verify that the suppressor is turned on and the current is set to the correct
value. Refer to the suppressor manual for additional troubleshooting
guidance. Suppressor manuals are included on the Dionex Reference Library
CD-ROM (P/N 053891).

Sample concentration too high

Dilute the sample (see

Section 3.9.3

).

Wrong eluent

Check that you are using the correct eluent for your application.

Background conductivity not offset from conductivity reading

Before making an injection, allow the background conductivity to equilibrate,
and then click

Autozero

on the ICS-2100Control panel in Chromeleon or the

ICS-2100 touch screen

HOME

page.

Cell out of calibration

Recalibrate the cell (see

Section 5.1.3

).