4 installing the new egc holder without a cr-tc, Or epm, Section 5.23.4 – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 204

Advertising
background image

ICS-2100 Ion Chromatography System

192

Doc. 065291-01 3/09

8. Turn the EGC holder right side up and set it in place on top of the

ICS-2100.

9. Slide the EPM mounting bracket and connected EPM onto the side of

the EGC holder.

10. Feed the following liquid lines through the tubing chase to the ICS-

2100 front component panel.

TO SRS/AES REGEN OUT

TO INJ VALVE IN - P/ELUENT OUT

TO PUMP/DAMPER

11. Connect the lines to their respective locations:

Connect the line labeled

TO SRS/AES REGEN OUT

to the

REGEN

OUT

port on the suppressor.

Connect the line labeled

TO INJ VALVE IN - P/ELUENT OUT

to the

EGC CO

3

Mixer inlet.

Connect the line labeled

TO PUMP/DAMPER

to the pulse damper.

12. Direct the

WASTE, GAS SEPARATOR

line from the top of the ICS-

2100 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see

Section B.10.1

).

13. Go on to

Section 5.23.5

.

5.23.4 Installing the New EGC Holder Without a CR-TC or EPM

Follow the instructions in this section if a CR-TC or EPM is not installed.

1. Set the new EGC holder in place on top of the ICS-2100.

2. Feed the following liquid lines through the tubing chase to the ICS-

2100 front component panel.

TO SRS/AES REGEN OUT

TO INJ VALVE IN - P/ELUENT OUT

TO PUMP/DAMPER

3. Connect the lines to their respective locations:

Connect the line labeled

TO SRS/AES REGEN OUT

to the

REGEN

OUT

port on the suppressor.