4.2

Troubleshooting Error Messages

This error occurs when the auxiliary power supply is configured as active, but
no device is connected.

To troubleshoot:

1. Connect the cable for the auxiliary device to the

EGC-2

power supply

connector.

2. If there is no auxiliary device, open the Chromeleon Server Configuration

program and remove this option from the ICS-2100 timebase.

This error occurs when the device connected to the auxiliary power supply is
attempting to draw more current than is allowed (200 mA) or is 20% higher
than the set current.

To troubleshoot:

1. Check the cable connecting the device to the power supply for damage or

a faulty connection.

2. If the error persists, refer to the manual provided by the device vendor for

assistance.

AUXILIARY POWER SUPPLY DISCONNECTED

AUXILIARY POWER SUPPLY OVER CURRENT