2 removing the cr-tc and reinstalling it, In the new egc holder, Section 5.23.2 – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 200

Advertising
background image

ICS-2100 Ion Chromatography System

188

Doc. 065291-01 3/09

5.23.2 Removing the CR-TC and Reinstalling it in the New EGC

Holder

1. Turn the holder upside down to view the CR-TC.

2. Grasp the top of the CR-

TC and lift up to
disconnect it from the
holder.

3. Disconnect the liquid

lines from the four ports
on the CR-TC (see

Figure 5-59

).

Figure 5-76. Disconnecting CR-TC Liquid Lines

4. Turn the new holder

upside down (see

Figure 5-77

) and

connect the liquid lines
to the CR-TC.

NOTE To help retain the

fittings and ferrules
on the lines, remove
the coupler from
each line just before
connecting the line
to a port.

Figure 5-77. EGC Holder (upside down)

Disconnect liquid
lines from 4 ports

Couplers

Metal stud for mounting the CR-TC