229

below 30%. At 0%, an error message informs you that the ion count is
depleted and the cartridge must be replaced before continuing
operation.

NOTE When the EGC-2 power supply is configured as

an auxiliary power supply, the EGC-2 Life
Remaining field is blank.

Normalized Life Used

(displayed when an EPM is installed)

indicates how much the EPM has been used. The value starts at 0 and
counts up with use. A count of 1 means the EPM usage is
approximately equal to the lifetime of one K

2

CO

3

cartridge.

B.8.5

Controlling the CR-TC Power

To manually control the CR-TC power, touch the

CR-TC Off

and

On

buttons on the

ELUENT GENERATOR

page (see

Figure B-17

).

To automatically control the CR-TC power simultaneously with the
pump flow, select

Auto turn on with pump

.

NOTE The CR-TC power always turns off when the pump

is off, regardless of whether the “Automatically turn
on with pump” option is selected.

Figure B-17. Eluent Generator Page: CR-TC Controls