ICS-2100 Ion Chromatography System

224

Doc. 065291-01 3/09

To change the low pressure limit, touch the

LOW

field and enter the

desired limit on the number pad.

To select the pressure unit for all touch screen pages that display pressure
values, touch the pressure unit field and select the desired unit on the list
(

PSI

,

bar

, or

MPa

) (see

Figure B-14

).

B.7.2

Setting Degas Operating Parameters (Optional)

By default, the ICS-2100 monitors the degas pressure reading and turns
the degas pump on and off as required. To change the degas operating
option:

1. Select

PUMP

from the menu of pages.

2. Under

Degas Settings

, touch the options drop-down list. A list of

available options appears (see

Figure B-15

).

NOTE Degas options can also be selected from the

Chromeleon Server Configuration program
(see

Section 2.4.1

).

Figure B-14. Pump Page: Selecting the Pressure Unit

Figure B-15. Pump Page: Selecting Degas Options

Select the
pressure unit

Select the
degas option