A.3 environmental, A.4 front panel, A.5 analytical pump and fluidics – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 208: Environmental, Front panel, Analytical pump and fluidics, Refer to, Section a.3

Advertising
background image

ICS-2100 Ion Chromatography System

196

Doc. 065291-01 3/09

A.3

Environmental

A.4

Front Panel

A.5

Analytical Pump and Fluidics

Operating

Temperature

4 to 40 °C (40 to 104 °F); cold room-compatible (4 °C) as long
as system power remains on

Operating

Humidity

5% to 95% relative humidity, noncondensing

Operating Pressure

35 MPa (5000 psi) maximum liquid path (tubing, valve,
columns, etc.)

Power LED

On when power is present; off when no power is present

LCD Touch

Screen

¼ VGA (320 x 240 pixels) liquid crystal display with touch screen,
which provides:

Local control of ICS-2100 components: pump, detector,
suppressor, injection valve, eluent generator, CR-TC, column
heater, vacuum degas, TTLs, and relays

Display of conductivity data plot and operating status

Local control of calibration and diagnostic functions

Type

Serial dual-reciprocating pistons, microprocessor-controlled
constant stroke, variable speed

Construction

Chemically inert, metal-free PEEK pump heads and flow paths
compatible with aqueous eluents of pH 0 to 14 and reversed-phase
solvents

Operating

Pressure

0 to 35 MPa (0 to 5000 psi)

Flow Rate

Range

0.00 to 5.00 mL/min in 0.01 mL/min increments without changing
pump heads

Flow Precision

<0.1% typical