2 replacing a k2co3 cartridge – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 163

Advertising
background image

5 • Service

Doc. 065291-01 3/09

151

11. Lift the EluGen

cartridge out of the
holder, turn it over,
and set it in the
service area.

12. Disconnect the

backpressure coil
from the EluGen
cartridge

OUTLET

port

.

13. Connect the

EGC

OUT

line from the

holder to the

OUTLET

port (see

Figure 5-45

).

Figure 5-45. EluGen Cartridge Inlet and Outlet Connections

14. Turn over the EluGen cartridge (fittings facing down). Check for

bubbles. If necessary, shake and tap the EluGen cartridge to remove
bubbles.

15. Reinstall the EluGen cartridge in the holder.

16. Reconnect the EluGen cartridge cable.

17. Reconnect the CR-TC cable.

18. If you disabled the

Automatically turn on with pump

option in

Step 3

, select the option.

5.19.2 Replacing a K

2

CO

3

Cartridge

1. Turn off the pump from the ICS-2100 Control panel in Chromeleon

or from the ICS-2100 touch screen

HOME

page.

2. Disconnect the EluGen cartridge cable from the

EGC-1

connector on

the ICS-2100 top cover.

3. If an EPM is installed, disconnect the EPM cable from the

EGC-2

connector.

EGC OUT

OUTLET

Port

EGC IN

INLET

Port