23 replacing the egc holder and degas assembly, 1 disconnecting and removing the elugen cartridge – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 197

Advertising
background image

5 • Service

Doc. 065291-01 3/09

185

17. Reconnect the

TO SRS/AES REGEN OUT

line to the suppressor

REGEN

OUT

port.

18. Reconnect the mixer outlet line to port

P

(

2

) on the injection valve.

5.23 Replacing the EGC Holder and Degas Assembly

5.23.1 Disconnecting and Removing the EluGen Cartridge

1. Turn off the power to the ICS-2100.

2. Disconnect the following EGC holder liquid lines:

Disconnect the line labeled

TO INJ VALVE IN - P/ELUENT OUT

from either port

P

(

2

) on the injection valve, or if an EGC CO

3

Mixer inlet is installed, from the mixer inlet.

If a CR-TC is installed, disconnect the line labeled

TO SRS/AES

REGEN OUT

from the

REGEN OUT

port on the suppressor.

Disconnect the line labeled

TO PUMP/DAMPER

from the pump

damper.

Disconnect the line labeled

WASTE-GAS SEPARATOR

from the

gas separator waste tube assembly installed on the waste
container.

3. Disconnect the EluGen cartridge cable, and the CR-TC or EPM cable

(if installed) from their respective connectors on the ICS-2100 top
cover.

The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.

La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.

Die EluGen-Kartusche enthält eine korrosive Base (KOH, LiOH oder
NaOH), oder eine korrosive Säure (MSA). Tragen Sie daher beim
Umgang mit der Kartusche eine Schutzbrille und Handschuhe.