1 loading samples with a syringe – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 75

Advertising
background image

3 • Operation and Maintenance

Doc. 065291-01 3/09

63

Setup for an AS-HV Autosampler

For setup information for the AS-HV Autosampler, see the AS-HV Autosampler
Operator’s Manual
(Document No. 065125). The manual is provided on the
Dionex Reference Library CD-ROM (P/N 053891).

3.10.1 Loading Samples with a Syringe

1. Make sure the injection port on the ICS-2100 front door (see

Figure 2-1

) is connected to sample port

S (5)

on the injection valve

(see

Figure 3-10

).

2. Fill the 1 cc syringe (P/N 016388) provided in the ICS-2100 Ship Kit

(P/N 064375) with a calibration standard or sample.

3. Insert the syringe into the injection port on the ICS-2100 front door

(see

Figure 2-1

).

4. Verify that the injection valve is in the Load position.

5. Overfill the sample loop with several sample loop volumes. Excess

sample will exit through the injection valve waste line.

6. Leave the syringe in the port.

7. Switch the injection valve to the Inject position (see

Section 3.10.4

).

Figure 3-10. Injection Valve Connections