Doc. 065291-01 3/09

249

Connect the Analog Output

1. Locate the analog cable for the third-party analog-to-digital converter and

controller.

2. Connect the analog input connector on the third-party controller to the

ANALOG OUTPUT

connector on the ICS-2100 rear panel (see

Figure C-6

).

Configure the ICS-2100 Analog Output

Go to the ICS-2100

DETECTOR

touch screen page and select the settings shown in

Figure C-9

:

Create an AS Method and Schedule

Figure C-10

is an example of AS method timed events that can be used with the

TTL/Relay connections shown in

Figure C-6

.

Figure C-9. Detector Page

Figure C-10. Example AS Method for Stand-Alone Operation

Help Prompt

METHOD:

TIMED EVENTS

1

1

1

TIME

INIT LOAD

INJECT

0.00

VALVE

TTL1 TTL2 RLY1 RLY2

1

CSV

0

0

1

0.50