5.21 Replacing the EPM Electrolytic pH Modifier

1. Turn off the pump from the ICS-2100 Control panel in Chromeleon or from

the touch screen

HOME

page.

2. Disconnect the EPM electrical cable from the

EGC-2

connector on the ICS-

2100 top cover.

3. Disconnect the four liquid lines from the ports on the EPM.

4. Lift up the mounting bracket with the EPM attached and remove the bracket

and EPM from the EGC holder.

5. Pull the EPM off the mounting bracket.

6. Remove the plugs from the

ports on the new EPM
(P/N 063175).

7. Push the EPM onto the ball

studs on the mounting bracket
(see

Figure 5-63

).

Figure 5-63. EPM and Mounting Bracket

8. Reinstall the mounting bracket with new EPM attached onto the EGC holder.

9. Reconnect the liquid lines that were disconnected in

Step 3

to the

corresponding ports on the new EPM.

10. Connect the EPM electrical cable to the

EGC-2

connector.

EGC Cable

Ball Studs

Mounting Bracket

EPM