B • Touch Screen Operation

Doc. 065291-01 3/09

223

B.7

Pump Page

Use the

PUMP

page to set parameters related to pump operation, including

pressure limits, pressure unit, vacuum degas pump settings, and eluent valve
control. To open this page, select

PUMP

from the menu of pages. The current

system pressure is displayed at the top of the screen.

B.7.1

Setting Pump Pressure Limits and Selecting the

Pressure

Unit

The first time the power to the ICS-2100 is turned on, the maximum
system pressure limit is 34.5 MPa (5000 psi) and the minimum pressure
limit is 0. The eluent generator, however, requires a maximum high limit
of 20.7 MPa (3000 psi) and a minimum low limit of 1.4 (200 psi). When
you configure the eluent generator cartridge in Chromeleon (see

Section B.8

), the software sets the maximum and minimum limits for the

pump automatically. If you are not using Chromeleon to control the ICS-
2100, set the limits on the

PUMP

page.

To change the high pressure limit, touch the

HIGH

field and enter the

desired limit on the number pad. The high limit must be at least
0.7 MPa (100 psi) above the low limit.

Figure B-13. ICS-2100 Touch Screen Pump Page

Current
system
pressure