4 • Troubleshooting

Doc. 065291-01 3/09

91

Flow rate is set to 0

Set the flow rate from the ICS-2100 Control panel in Chromeleon or from the
touch screen

HOME

page.

Eluent valve is closed

Open the eluent valve from the ICS-2100 Control panel in Chromeleon or
from the touch screen

PUMP

page.

Pump not primed

Prime the pump (see

Section 5.17

).

Broken pump piston

Replace the piston (see

Section 5.8

).

Eluent generator degas tubing assembly ruptured

Replace the EluGen cartridge holder (see

Section 5.19

).

4.7

Erratic Flow/Pressure Reading

Pump needs priming

Prime the pump (see

Section 5.17

).

Figure 4-2. Waste and Priming Valves

Waste Valve

Priming Valve