5 • Service

Doc. 065291-01 3/09

137

12. Align the new valve on the component panel with the liquid ports facing up

and attach it with the two screws removed in

Step 5

. Reconnect the liquid

lines.

13. Turn on the power to the ICS-2100.

14. Prime the pump (see

Section 5.17

).

5.15 Replacing the Leak Sensor

1. Turn off the power to the ICS-

2100.

2. Open the ICS-2100 front door.

3. Loosen the screw on the front of

the leak sensor (see

Figure 5-29

). Note: The screw

remains attached to the sensor.

Figure 5-29. Leak Sensor

4. Remove the leak sensor from

the component panel and pull
the cable out of the opening in
the panel.

5. The cable attached to the leak

sensor is connected to a
matching cable inside the ICS-
2100. Continue pulling the cable
until the connectors for the two
cables are outside of the panel
(see

Figure 5-30

).

6. Disconnect the two cables.

Figure 5-30. Leak Sensor Cable

Leak

Sensor

Loosen
Screw

Disconnect

Cables