B • Touch Screen Operation

Doc. 065291-01 3/09

221

2. Touch the down arrow next to the field you want to change. A list of

the available status parameters appears (see

Figure B-12

).

3. Touch the parameter you want to display.

4. When finished, touch the

Apply

button.

B.6.2

Status Parameter Details

Figure B-12. Status Page Parameter List

Status Parameter

Description

Eluent time

The number of hours the eluent supply will last, if the pump runs
continuously at the current flow rate.

Flow rate

The actual flow rate currently output by the pump.

EGC1 conc
EGC2 conc

The actual concentration of eluent currently being generated by the
EluGen cartridge (EGC1 or EGC2).

AUX PS

The actual auxiliary power supply current in mA.
This field replaces the

EGC2 conc

field when the EGC-2 power

supply is configured as an auxiliary power supply.

Suppressor

The actual current applied to the suppressor.

Elapsed time

The number of hours since the last injection.