5 • Service

Doc. 065291-01 3/09

125

5.9

Replacing the Waste Valve or Priming Valve O-Ring

A damaged O-ring causes leakage around the base of the waste valve or
priming valve knob.

1. Close the eluent valve from the ICS-2100 Control panel in Chromeleon or

select

Eluent Valve Closed

on the touch screen

PUMP

page.

2. Turn off the main power switch, to ensure that you do not unintentionally start

the ICS-2100.

3. To remove the waste valve or priming valve from the pump head (see

Figure 5-16

), turn the knob counterclockwise until it is loose, and then pull

the knob straight out of the cavity in the pump head.

Figure 5-16. Waste and Priming Valves

Waste Valve

Priming Valve