ICS-2100 Ion Chromatography System

108

Doc. 065291-01 3/09

5.1.5

Calibrating the Vacuum Degas Assembly

When to Calibrate

After every 6 months of use

1. Open the ICS-2100 Wellness Panel in Chromeleon (see

Section 5.1.1

), or open the ICS-2100 touch screen

PUMP DEGAS

page

(see

Figure 5-6

).

2. On the Wellness Panel under Degas Calibration, select Calibrate, or

touch

Calibrate

on the ICS-2100 touch screen. The degas pump runs

for 90 seconds to allow a vacuum to be created. After 90 seconds, the
vacuum degas pressure reading is recorded. When the vacuum degas
assembly is running in monitor mode, this calibration value is used to
determine when to turn on the degas pump.

3. If you are using the Wellness panel, click

Log

to record the new

calibration value in the Audit Trail.

4. If you performed the calibration from the ICS-2100 touch screen,

record the new calibration values in the Chromeleon Wellness
database. To do this, open the Wellness Panel (see

Section 5.1.1

) and

click Upload under Update Wellness Database.