5.21.1 Recording the EPM Serial Number in Chromeleon

The serial number is printed on a label on the EPM.

1. Open the Server Configuration program (select Start > All

Programs > Chromeleon > Server Configuration).

2. Double-click the ICS-2100 system icon. The properties dialog box for

the ICS-2100 appears.

3. Click the Eluent Generator tab (see

Figure 5-44

).

4. Select the EGC-2 check box.

5. Enter the serial number of the EPM.

6. Verify that the cartridge type is

pH Mod

. If not, make sure the serial

number was entered correctly.

7. Click OK.

Figure 5-64. Eluent Generator Configuration Properties Example