ICS-2100 Ion Chromatography System

212

Doc. 065291-01 3/09

B.4

Home Page

B.4.1

Home Page Pump Controls

Pressure Display

The

HOME

page indicates the current pressure reading from

the pressure transducer. The pressure unit can be set to psi
(the factory default), MPa, or bar (see

Section B.7.1

).

The minimum and maximum pressure limits are displayed
below the current pressure reading. If the pressure reading
goes outside these limits, the pump is stopped and an error
message appears in the Chromeleon Audit Trail and on the
touch screen. See

Section 4.2

for troubleshooting

information. See

Section B.7.1

for how to set the pressure

limits.

Figure B-7. ICS-2100 Touch Screen

Home

Page

Edit Field

Command

Button

(selected)

Command

Button

(unselected)

Status Field