B.4.2, B.4.3 – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 226

Advertising
background image

ICS-2100 Ion Chromatography System

214

Doc. 065291-01 3/09

B.4.2

Home Page EGC Controls

The

EGC

controls indicate the

type and the eluent
concentration of each installed
EluGen Cartridge (EGC). The

CR-TC

field indicates whether

the Continuously Regenerated
Trap Column is on or off.

To determine the eluent concentration for your application, refer to the column
manual. Column manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891). When the ICS-2100 is controlled from the front panel, only
isocratic eluent delivery is possible. With isocratic delivery, the eluent
composition and concentration remain constant throughout the run. When the
ICS-2100 is under Chromeleon control, gradient delivery, in which the eluent
concentration changes over time, is also possible.

To set the eluent concentration or to turn on or off the CR-TC, touch the
concentration button. The

ELUENT GENERATOR

page appears (see

Section B.8

).

NOTE When the EGC-2 is configured as an auxiliary power

supply, the EGC-2 cartridge type field displays

AUX

and

the EGC-2 eluent concentration field displays the actual
current (in mA) of the auxiliary power supply.

B.4.3

Home Page Column Heater

The temperature of the column heater can be controlled to
between 30 and 60 °C. For optimal performance, set the
column heater to at least 5 °C above the ambient
temperature. To set the temperature, touch the

Column

Heater

field and enter the desired temperature on the number pad. To turn the

column heater off, set the temperature to 0. The display to the right of the
temperature setting indicates the heater status:

=

column temperature is at equilibrium

column temperature is rising

column temperature is falling

Off

column heater is off

EGC-1 Eluent
Concentration

EGC-1

Cartridge

Type

EGC-2

Cartridge

Type

EGC-2 Eluent
Concentration