2 • Features

Doc. 065291-01 3/09

39

details, refer to the EluGen cartridge manual. Cartridge manuals are
included on the Dionex Reference Library CD-ROM (P/N 053891).

EluGen Cartridge Holder

The EluGen cartridge is installed in a cartridge holder, which mounts on
the ICS-2100 top cover (see

Figure 2-7

). The cartridge holder also houses

the high-pressure degas tubing assembly.

The ICS-2100 can control up to two EluGen cartridges. However, when
operating with an EGC II K

2

CO

3

EluGen Cartridge and the pH Modifier,

only one cartridge can be installed, because the pH Modifier must be
connected to the other EGC power supply (

EGC-1

or

EGC-2

).

EluGen
Cartridges

Eluent Concentration Range

Comment

K

2

CO

3

/EPM

Electrolytic
pH Modifier

0.1 to 15 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 2.0 mL/min
flow
X = 15/flow

The total of the eluent
concentrations from
both the EluGen
cartridge and the EPM
must not exceed the
specified range. The
EPM concentration
must not exceed
10 mM. See

“Generating
Carbonate/Bicarbonate
Eluent” on page 41

for

details.

KOH/KOH
KOH/MSA
KOH/NaOH
MSA/MSA
MSA/NaOH
NaOH/NaOH

0.1 to 50 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 3.0 mL/min
flow
X = 50/flow

The eluent
concentration range for
each cartridge is cut by
50%.

LiOH/LiOH

0.1 to 40 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 3.0 mL/min
flow
X = 40/flow

The eluent
concentration range for
each cartridge is cut by
50%.

Table 2-3. Eluent Concentration Ranges for Linked

Dual-Cartridge Configurations