Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 116

Advertising
background image

ICS-2100 Ion Chromatography System

104

Doc. 065291-01 3/09

2. On the Wellness Panel under Electric Conductivity Cell

Calibration, click Offset Cal, or touch

Run Offset Cal

on the ICS-

2100 touch screen.

3. When the offset calibration is complete, click Slope Cal on the

Wellness Panel, or touch

Run Slope Cal

on the ICS-2100 touch

screen.

4. If you are using the Wellness Panel, click Log to record the new

calibration values in the Audit Trail.

5. Disconnect the pump output line from the injection valve.

6. Disconnect the line from the suppressor

ELUENT OUT

port to the cell

inlet and connect the pump output line directly to the cell inlet.

7. Disconnect the line from the suppressor

REGEN IN

port to the cell

outlet. Direct the cell outlet line to a waste container.

8. Verify that the backpressure is at least 7 MPa (1000 psi).

9. If you are using the touch screens to perform the calibration, open the

CONDUCTIVITY CELL

page (see

Figure 5-4

).

10. On the Wellness Panel under Conductivity Cell Calibration, click

Cell 35 °C, or touch

35 C

on the ICS-2100 touch screen. Allow the

cell to reach this temperature, and then wait an additional 5 minutes to

Figure 5-4. Conductivity Cell Page