ICS-2100 Ion Chromatography System

218

Doc. 065291-01 3/09

B.6

Status Page

The

STATUS

page displays the current status of various operating parameters. To

open this page, touch the

STATUS

button on the

HOME

page or select

STATUS

from

the menu of pages.

Status Page Controls

Pressure and conductivity readings are the same as those displayed on the

HOME

page.

The eluent bottle depicts the level of liquid in the eluent reservoir. The scale is
always 0 to 4 liters (regardless of the size of reservoir installed). After filling
the reservoir, go to the

HOME

page and enter the amount of liquid in the

reservoir (see

Section B.4.1

). The ICS-2100 monitors the eluent usage, based

on the flow rate and the length of time the pump is on. The “liquid” level in
the bottle falls as the reservoir empties. An error message is first displayed on
the screen when the level falls below 200 mL. The message is repeated at
100 mL and then at 0 mL.

Figure B-9. Status Page

In order for the eluent level displayed on the screen to be accurate,
the level must be entered by the user when the reservoir is filled. The
ICS-2100 does not automatically detect when the reservoir is filled,
nor when it is empty.