Isolating a Restriction in the Liquid Lines

A restriction in the liquid plumbing will cause excessive system backpressure.

1. Begin pumping eluent through the system (including the columns).

2. Follow the flow schematics in

Figure 5-7

,

Figure 5-8

, or

Figure 5-9

and work

backward through the system, beginning at the suppressor

Regen Out

port.

One at a time, loosen each fitting and observe the pressure. The connection at
which the pressure drops abnormally indicates the point of restriction.

NOTE The numbers on the flow schematics indicate the order

in which liquid flows through the system components.

If the restriction has caused such high pressure that the system cannot be
operated, you must work forward through the flow schematic, adding parts
one at a time until an abnormal pressure increase (and hence, the restriction) is
found.

3. If the restriction is in the tubing or fitting, remove the restriction by back

flushing or by replacing the tubing or fitting (see

Section 5.3

).