B.4.4, B.4.5 – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 227

Advertising
background image

B • Touch Screen Operation

Doc. 065291-01 3/09

215

B.4.4

Home Page Injection Valve Controls

The

Inject

and

Load

buttons select the position of the

injection valve. A dark background and white text on a
button indicates the current valve position. For details about
the valve positions, see

Section 2.4.6

.

The

ELAPSED TIME

field displays the amount of time since

the last injection.

B.4.5

Home Page Detector Controls

Conductivity Readings

The offset conductivity is displayed in large numbers above
the total conductivity. The offset conductivity is the total
conductivity minus the current offset.

To determine the offset, allow the system to equilibrate after

startup. At equilibration, the conductivity reading is the background conductivity
(the eluent conductivity before sample injection). Touch

Autozero

to zero the

conductivity and offset this background reading from the total reading.

NOTE You can reverse the conductivity display to show the

total conductivity in large type, with the offset
conductivity below it in small type. Select this option on
the

MODULE SETUP

page (see

Section B.12

).

Cell Heater Controls

The temperature of the cell can be set to between 30 °C and
55 °C, but must be at least 7 °C above the ambient
temperature. To set the temperature, touch the

Cell Heater

field and enter the desired temperature on the number pad.

To turn off the cell heater, set the temperature to 0. The display to the right of the
temperature setting indicates the heater status:

=

cell temperature is at equilibrium

cell temperature is rising

cell temperature is falling

Off

cell heater is off