C • TTL and Relay Control

Doc. 065291-01 3/09

241

b. Plug the connector into the 12-pin connector on the ICS-2100 rear panel.

c.

Connect the wires from the ICS-2100 connector plug to the TTL or relay
connector pins on the other module(s). Additional connector plugs are
provided with other Dionex modules.

NOTE Check the polarity of each connection. Connect

signal wires to signal (+) pins and ground wires to
ground (-) pins.

3. If you connected a TTL input, verify that the correct function is assigned to

the input and that the correct input control type is selected. Select different
settings, if necessary. Input functions and control types are assigned from
either the

INPUT/OUTPUT

touch screen page or Chromeleon (see

Section C.1.1

).