2 filling the egc co3 mixer with deionized water, 5 • service, 2 filling the egc co – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 187: Mixer with deionized water

Advertising
background image

5 • Service

Doc. 065291-01 3/09

175

4. Connect the

TO EGC CO3 MIXER IN

line to the mixer inlet and connect

the

TO EGC CO3 MIXER OUT

line to the mixer outlet.

5.22.2 Filling the EGC CO

3

Mixer with Deionized Water

Before initial use, follow the steps below to temporarily fill the EGC CO

3

Mixer with deionized water. Then, follow the steps in the next section to
fill the mixer with eluent of the required concentration for the application
to be run.

1. Verify that the EGC-1 and EGC-2 currents are off.

2. Disconnect the EluGen cartridge inlet line (labeled

TO

PUMP/DAMPER

) from the coupler that connects it to the green pump

outlet line (see

Figure 5-68

).

Figure 5-67. ICS-2100 with an EGC CO

3

Mixer Installed

TO EGC CO

3

MIXER IN

TO EGC CO

3

MIXER OUT