ICS-2100 Ion Chromatography System

174

Doc. 065291-01 3/09

14. Reconnect the

TO INJ VALVE IN - P/ELUENT OUT

line to the EGC CO

3

Mixer inlet.

15. Reconnect the

TO SRS/AES REGEN OUT

line to the suppressor

REGEN

OUT

port.

5.22 Replacing the EGC CO

3

Mixer

5.22.1 Installing the New EGC CO

3

Mixer

1. Turn off the pump from the ICS-2100 Control panel in Chromeleon

or from the touch screen

HOME

page.

2. Disconnect the inlet and outlet lines (labeled

TO EGC CO3 MIXER IN

and

TO EGC CO3 MIXER OUT

) from the existing mixer and pull the

mixer off the mounting clips.

3. Orient the new EGC CO

3

Mixer (P/N 061686)

with the outlet

pointing upward and push it onto the mounting clips (see

Figure 5-67

). The mixer outlet must point upward to ensure thorough

mixing of the eluent.