B • Touch Screen Operation

Doc. 065291-01 3/09

225

3. Select the desired option:

Always Off

: The degas pump is always off.

Always On

: The degas pump is always on. This setting is reserved for

test purposes by a Dionex Service Representative.

Duty Cycle

: The degas pump cycles on and off according to the times

specified in the

Duty Cycle

and

Time Off

fields.

Duty Cycle

specifies

for how long the degas pump runs during a cycle;

Time Off

specifies

the time between cycles.

Monitor

: The ICS-2100 monitors the degas pressure reading and

turns the degas pump on and off as required.

B.7.3

Controlling the Eluent Valve

The eluent valve controls flow from the
eluent reservoir. During routine operation,
the valve opens and closes automatically

when the pump flow is turned on and off. However, during eluent line
priming, you can use the controls on the

PUMP

page to manually open the

valve, allowing liquid to flow through the lines when the pump is off. See

Section B.13.2

for eluent line priming instructions.

Select the Degas, Always On option only when testing the vacuum degas
assembly. Do not select the Always On option for routine use.