4.10 Nonreproducible Peak Height or Retention Time

Column overloading

Dilute the sample (see

Section 3.9.3

).

Liquid leaks

Locate and eliminate the leaks (see

Section 4.3

).

Incomplete or imprecise filling of the sample loop

1. Fill the sample loop until excess sample exits the waste line.

2. Inspect the 1-cc syringe (P/N 016388) and replace if damaged.

Pump not primed properly

Prime the pump (see

Section 5.17

).

4.11 Abnormal Retention Time or Selectivity

Incorrect eluent

Check the EGC concentration setting.

Contaminated or degraded sample

Take appropriate precautions when preparing and storing samples to prevent
contamination and degradation (see

Section 3.9

).

Contaminated column

1. Clean the column as instructed in the column manual. Column manuals

are included on the Dionex Reference Library CD-ROM (P/N 053891).

2. If cleaning is unsuccessful, replace the column.

4.12 No Cell Response

Cell not properly installed

Verify that the cell front cover is flush against the component panel. If
necessary, tighten the two mounting screws. When the cell is correctly
installed and the screws tightened, an electronics connector on the cell plugs
into a receptacle inside the ICS-2100.