A.14 suppressors, A • specifications – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 215

Advertising
background image

A • Specifications

Doc. 065291-01 3/09

203

A.14 Suppressors

Chemical

Suppression

2 mm and 4 mm anion and cation, membrane suppression bed types

Displacement

Chemical

Regeneration

2 mm and 4 mm anion and cation, membrane suppression bed types

Electrolytic

Suppression

Self-Regenerating: 2 mm and 4 mm anion and cation; both
membrane and MonoDisk

suppression bed types available

Self-Regenerating with External Water Mode: 2 mm and
4 mm anion and cation; both membrane and MonoDisk
suppression bed types available

Current Control

Range

SRS (4 mm): 0 to 300 mA in 1 mA increments
SRS (2 mm): 0 to 100 mA in 1 mA increments
AES: 0 to 150 mA in 1 mA increments
CMD: 0 to 500 mA in 1 mA increments
SRN: 0 to 500 mA in 1 mA increments

Salt Converter

2 mm and 4 mm versions available

AMMS ICE

®

2 mm and 4 mm versions available

Carbonic Acid

Removal for

Anions

ASRS

®

300 with CRD 200 for hydroxide eluents

ASRS 300 with CRD 300 for carbonate eluents

Non-Suppressed

Supported

Suppressor Wear

Parts

None; peristaltic pump and inline filters not required

Suppression

Capacity

Anion SRS 300 (4 mm): 200

μeq/min

Cation SRS 300 (4 mm): 110

μeq/min

Anion SRS 300 (2 mm): 50

μeq/min

Cation SRS 300 (2 mm): 37.5

μeq/min

Anion MMS 300 (4 mm): 150

μeq/min

Cation MMS 300 (4 mm): 150

μeq/min

Anion MMS 300 (2 mm): 37.5

μeq/min

Cation MMS 300 (2 mm): 37.5

μeq/min

Anion AES: 25

μeq/min

Cation AES: 25

μeq/min

Advertising