Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 141

Advertising
background image

5 • Service

Doc. 065291-01 3/09

129

To calibrate the cell on the ICS-2100 touch screen

DIAGNOSTIC

page:

a.

Go to the

DIAGNOSTIC

page and touch

Electric Conductivity Cell

.

b. On the

ELECTRIC CONDUCTIVITY CALIBRATION

page (see

Figure 5-20

),

touch

Run Offset Cal

.

c.

When the offset calibration is complete, touch

Run Slope Cal

.

d. Touch the

Cell Constant

field and enter the value printed on the front of

the cell.

e.

To ensure that the new calibration values are recorded in the ICS-2100
memory, do not turn off the ICS-2100 power for at least 1 minute.

f.

Record the new calibration values in the Chromeleon Wellness database.
To do this, open the Wellness Panel (see

Section 5.1.1

) and click Upload

under Update Wellness Database.

Figure 5-20. Electric Conductivity Calibration Page