19 replacing an elugen cartridge, 1 replacing a koh, lioh, msa, or naoh, Section 5.19 – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 155

Advertising
background image

5 • Service

Doc. 065291-01 3/09

143

3. Replace the two fuses in the holder with new IEC 127 fast-blow fuses rated

3.15 amps (P/N 954745). Dionex recommends always replacing both fuses.

4. Reinsert the fuse holder into its compartment and push in until the tab clicks

in place.

5. Reconnect the main power cord and turn on the power.

5.19 Replacing an EluGen Cartridge

The procedure for replacing an EluGen cartridge differs, depending on the type of
cartridge:

To replace a KOH, LiOH, MSA, or NaOH EluGen cartridge, go to

Section 5.19.1

.

To replace a K

2

CO

3

cartridge, go to

Section 5.19.2

.

5.19.1 Replacing a KOH, LiOH, MSA, or NaOH EluGen Cartridge

1. Turn off the pump from the ICS-2100 Control panel in Chromeleon

or from the ICS-2100 touch screen

HOME

page.

2. Disconnect the EluGen cartridge cable from the

EGC-1

connector on

the ICS-2100 top cover.

3. Disconnect the CR-TC cable from the

CR-TC

connector.

The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.

La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.

Die EluGen-Kartusche enthält eine korrosive Base (KOH, LiOH oder
NaOH), oder eine korrosive Säure (MSA). Tragen Sie daher beim
Umgang mit der Kartusche eine Schutzbrille und Handschuhe.